The tender for steam indicator tap class i , steam indicator tap class 5 moving front technology , steam biological indicators srbi , eto cartridge 100 gram big eto , eto cartridge 40 gram small eto , eto bio logical indicators srbi , bowie dick pack lead free , ultra rapid multienzyme cleaner 1 liter , fabric incise drape iodinised , chg 4 per scrub 500ml , skin prep chg 2per solution and 70per isopropenol alcohol in 500ml , eto packing roll size 100 200 , eto packing roll size 150 200 , eto packing roll size 250 200 , eto packing roll size 400 200, boq cssd, lead free autoclave tape roll, self adhesive moving front chemical integrators indicating accept or
reject to verify all perimeters of steam sterilization together i e time temperature and quality of steam
can be used for pack control monitoring of all 118 138c 245 280f steam sterilization cycles
should be compliant to bs en iso 11140 1 2005
should have 510 k registration
should have a bsi kitemark certificate pack of 500 pc 2 box case, rapid readout biological indicators for steam sterilization containing a
standardized viable population of bacillus stearothermophilus spores
the indicator detects the presence of viable spores by the production of fluorescence in 3 hours in conjunction with rapid readout incubator
designed and certified as per iso 11138 and en 866
should be compatible with the following cycles
a) 121 123c gravity steam sterilization cycle
b) 132 135c vacuum assisted steam sterilization cycles should be compliant to iso 11138 1 2006 en part 1 and iso 11138
1 2006 en part 3
every box should contain quality assurance certificate with mention of spore population d value z value
should have a bsi kitemark certificate
should have fda clearance, for eto sterilization, rapid readout biological indicators containing a standardized viable population
of bacillus atropheus spores for eto sterilization cycles
the indicator detects the presence of viable spores by the production of fluorescence within 4 hours in conjunction with rapid readout incubator
designed and certified as per iso 11138 and en 866 should be compliant to iso 11138 1 2006 en part 1 and iso 11138
1 2006en part 2
every box should contain quality assurance certificate with mention of spore populationd value z value
should have a bsi kitemark certificate
should have fda clearance, iso 11140 1 2014 and 11140 5 2007
designed for testing air-removal efficiency of 132 134 c 270 273 f dynamic air removal steam sterilizers
test pack will reproducibly detect residual air in the sterilizer chamber due to air leaks, air removal failures and inadequate vacuum by testing the air removal efficiency of 132 134 c 270 273f dynamic air removal steam sterilizers
indicator sheet will turn a uniform dark color except when air removal failures such as air leaks occur
lead free
equivalent in performance to the bowie dick towel pack described in ansi aami st79
does not change visually 24 months when properly stored includes a scanable 2d bar code with lot number and expiration date for record keeping
includes a process indicator on the test pack label preassembled, single use designed for consistent performance results additional sponge sheet inside the pack for more challenge
ink pattern design for easy interpretation two year shelf life
sheet can be filed for record keeping, multiple enzymatic concentrate solution with lipases cellulases proteases and amylases
must act against all possible body secretions compatible to all instruments
fast acting within 2 10 minutes allow safe disposable via the drain after use
leaves no residue after wash stable formulation with near neutral ph
should have scale inhibition properties
should help retain shine of the instrument
should be usable for manual ultra sonic and automatic cleaning systems
properties should have been tested by methods prescribed by nace tm0374 2007 nace tm0169 astm g31 2012 1 liter, incisde drape made up of polyester film with an acrylic adhesive that contains a complex of iodoform n vinyl 2 pyrrolidine with a iodine concentration of 0 092 mg iodine cm2 film should be breathable and conformable to skin compatible with povidone and chg skin preps iodine from the drape should be present in the deepest skin layers 1000 microns depth where hair folices exist apart from killing microbes on skin layer size 56 45 cm with frame delivery mechanism which allow application by single person, incisde drape made up of polyester film with an acrylic adhesive that contains a complex of iodoformn vinyl 2 pyrrolidine with a iodine concentration of 0 092 mg iodine cm2 film should be breathable and conformable to skincompatible with povidone and chg skin preps iodine from the drape should be present in the deepest skin layers 1000 microns depth where hair folices exist apart from killing microbes on skin layer size 34 35 cm with frame delivery mechanism which allow application by single person, pre operative surgical hand scrubbing washing by 4 per chlorhexidine solution can be used for patient s skin 500ml scrubbing cleansing and also for body washing showering should have moisturizer emollients to prevent skin damgeage dryness and irritation should provide higher foaming and grater viscocity to avoid wastage should have long persistent action broad spectrum persistentand cumulative antimicrobial activity slotution chlorhexidine gluconate ip, chlorhexidine gluconate scrub solution 2 per w v equivalent to 10per v v broad spectrum bactericidal fungicidal virucidal and sporicidal antimicrobial solution for skin preparation in a translucent bottle with volume calibration 500ml bottle from Vadodara, Gujarat may be relevant for contractors, suppliers, vendors, manufacturers, consultants, service providers, startups, MSMEs, and other organizations involved in chlorhexidine gluconate,biological indicators,pyrrolidine,calibration,cartridge,incubator,chemical,cleaning,bottle related products, services, or works. However, eligibility to participate depends on the qualification criteria specified by Gujarat medical education and research society GMERS in the official tender documents.
Before submitting a bid, interested applicants should carefully review the eligibility conditions, technical requirements, financial criteria, work experience requirements, annual turnover conditions, registrations, certifications, and any other supporting documents required by the tendering authority. The detailed qualification requirements and bidder eligibility criteria are available in the tender notice and related documents provided on this page.