The corrigendum tender for steam indicator tap class i , steam indicator tap class 5 moving front technology , steam biological indicators srbi , eto cartridge 10 gram big eto , eto cartridge 40 gram small eto , eto bio logical indicators srbi , dimethyl benzyl ammonium chloride solution dimethyl ethylbenzyl ammonium chloride solution , eto packing roll size 150 x 200 , eto packing roll size 250 x 200 , eto packing roll size 400 x 200 , bowie dick pack lead free , hydrogen peroxide and silver solution hp10 , ultra rapid multienzyme cleaner 1 liter , fabric incise drape iodinised , negative pressure wound therapy system small , negative pressure wound therapy system medium , negative pressure wound therapy system large, reagents and consumables of cssd, lead free autoclave tape roll, self adhesive ,moving front chemical integrators, indicating accept or
reject to verify all
perimeters of steam sterilization together i e time, temperature and quality of steam
can be used for pack control monitoring of all 118138 c 245
280 f steam sterilization cycles
should be compliant to bs en iso 111402005
should have 510 k registration
should have a bsi kitemark certificate pack of 500 pc 2 box
case, rapid readout biological indicators for steam sterilization containing a
standardized viable population of bacillus stearothermophilus spores
the indicator detects the presence of viable spores by the production of fluorescence in 3 hours in conjunction with rapid readout incubator
designed and certified as per iso 11138 and en 866
should be compatible with the following cycles
a 121 123 c gravity steam sterilization cycle
b 132 135 c vacuum-assisted steam sterilization cycles should be compliant to iso 111381 2006 en part 1 and iso 11138
1 2006en part 3every box should contain quality assurance certificate with mention of spore population d value z value
should have a bsi kitemark certificate
should have fda clearance, for eto sterilization, rapid readout biological indicators containing a standardizeviable population
of bacillus atropheus spores for eto sterilization cycles
the indicator detects the presence of viable spores by the production of fluorescence within 4 hours in conjunction with rapid readout incubator
designed and certified as per iso 11138 and en 866 should be compliant to iso 11138 1 2006 en part 1 and iso 11138
12006en part 2every box should contain quality assurance certificate with mention of spore population, d value z value
should have a bsi kitemark certificate
should have fda clearance, iso 1114012014 and 1114052007
designed for testing air removal efficiency of 132134c 270 273f dynamic-air removal steam sterilizers
test pack will reproducibly detect residual air in the sterilizer chamber due to air leaks, air removal failures and inadequate vacuum by testing the air removal efficiency of 132 134c 270 273 f dynamic air-removal steam sterilizers
indicator sheet will turn a uniform dark color except when air removal failures, such as air leaks, occur
lead free
equivalent in performance to the bowie-dick towel pack described in ansi aami st79
does not change visually 24 months when properly stored includes a scanable 2d bar code with lot number and expiration date for record keeping
includes a process indicator on the test pack label preassembled, single use designed for consistent performance results
ink pattern design for easy interpretation two year shelf life
sheet can be filed for record keeping, surface disinfectant, surface disinfectant and fogger solution, multiple enzymatic concentrate solution with lipases cellulases, proteases and amylases
must act against all possible body secretions, compatible to all instruments
fast acting within 2 10 minutes, allow safe disposable via the drain after use
leaves no residue after wash, stable formulation with near neutral ph.
should have scale inhibition properties
should help retain shine of the instrument
should be usable for manual, ultra sonic and automatic cleaning systems
properties should have been tested by methods prescribed by nace tm0374 2007 nace tm0169 astm g31 2012 1 liter, incisde drape made up of polyester film with an acrylic adhesive that contains a complex of iodoformn vinyl 2 pyrrolidine with a iodine concentration of 0.092 mg iodinecm2. film should be breathable and conformable to skin, compatible with povidone and chg skin preps iodine from the drape should be present in the deepest skin layers 1000 microns depthwhere hair folices exist apart from killing microbes on skin layer. size 26x20 cm with frame delivery mechanism which allow application by single person., negative pressure wound therapy system dressing kits containing small polyurethane ether black foam dressing10 x 7.5 x 3 cm with one drape, ruler, trac pad with double tubing along with 600 ml canister. it should be cdsco, iso ce certified should be eto sterilize, negative pressure wound therapy system dressing kits containing medium polyurethane ether black foam dressing 18 x 12 5 x 3 cm with one drape, ruler, trac pad with double tubing along with 600 ml canister. it should be cdsco, iso ce certified should be eto sterilize, negative pressure wound therapy system dressing kits containing large polyurethane ether black foam dressing 25 cm x 18 cm x 3 cm with one drape, ruler, trac pad with double tubing along with 600 ml canister. it should be cdsco, iso ce certified should be eto sterilize from Vadodara, Gujarat may be relevant for contractors, suppliers, vendors, manufacturers, consultants, service providers, startups, MSMEs, and other organizations involved in related products, services, or works. However, eligibility to participate depends on the qualification criteria specified by Gujarat medical education and research society GMERS in the official tender documents.
Before submitting a bid, interested applicants should carefully review the eligibility conditions, technical requirements, financial criteria, work experience requirements, annual turnover conditions, registrations, certifications, and any other supporting documents required by the tendering authority. The detailed qualification requirements and bidder eligibility criteria are available in the tender notice and related documents provided on this page.
Ensuring compliance with all eligibility and qualification requirements before bid submission can help bidders avoid disqualification and improve their chances of successful participation in the tender process.
Tende18 provides comprehensive tender bidding services, including a complete tender summary along with a detailed analysis of the Tender Qualification and Eligibility Criteria.